Even Ohm Leads and Even Ohm Heater Wire
The result of measurement for a thermal converter manufactured by Precision Measurements with Evanohm leads and Evanohm heater wire, thereby eliminating the Seebeck and Peltier effects, which exist in most cases where the heater wires and lead wires are dissimilar materials.
Applied voltage= .5 Volts
Measurements made using FRDCThe result of measurement for a thermal converter
manufactured with Evanohm leads and Evanohm heater wire, thereby eliminating
the Seebeck and Peltier effects,
which exist in most cases where the heater wires and lead wires are dissimilar
materials.
Applied voltage= .5 Volts
Measurements made using FRDC
Frequency | AC/DC (ppm) | Standard Deviation |
1.00E+01(Hz) | -.38 | +/-0.28 |
4.00E+01(Hz) | -.47 | +/-0.41 |
1.00E+02(Hz) | -.64 | +/-0.46 |
4.00E+02(Hz) | -.61 | +/-0.44 |
1.00E+03(Hz) | -.32 | +/-.35 |
4.00E+03(Hz) | .02 | +/-0.34 |
1.00E+04(Hz) | .09 | +/-0.35 |